Part No: VF-P32-A-01
Thermo Scientific Vanquish UHPLC pumps offer more performance without any tradeoff on durability and robustness, for highest system up-time and lowest total cost of ownership. The industry-leading Thermo Scientific SmartFlow pumping technology of the Vanquish pumps always provides you with unmatched retention time, reproducibility, and lowest baseline noise for highest detection sensitivity, independent of eluent composition and for backpressures up to 150 MPa (1500 bar, 22,000 psi) the productivity can be maximized using two pumps with Vanquish Duo HPLC and UHPLC workflows.
Min. Flow Rate | 0.001 mL/min. |
Operating Principle | Serial dual-piston pumps |
Solvent Degassing | Built-in, 6 Channels |
Biocompatible Version | Yes; pH Range: 2 to 12, Chloride Concentration: Up to 1 mol/L |
Flow Accuracy | ±0.1% |
Gradient Formation | Dual Low-Pressure Gradient Proportioning |
Max. Flow Rate | 8 mL/min. |
Type | Serial Dual-Piston Pump |
GLP Features | Predictive performance functions for scheduling maintenance procedures based on the actual operating and usage conditions of the pump. All system parameters logged in the Chromeleon CDS Audit Trail. |
Flow Precision | <0.05% RSD or <0.01 min SD, whichever is greater |
Pressure Range | 2 to 103 Mpa |
No. of Solvents | 2×3 |
Proportioning Accuracy | ± 0.5% of full-scale ± 1.0% (of full-scale) for other combinations than AB |
Proportioning Precision | < 0.15% SD |
PC Connection | USB 2.0, 3-port-HUB to connect other Vanquish modules |
Safety Features | Leak detection and safe leak handling, excess pressure monitoring |
Pressure Ripple | Typically: <1.0% or <0.2 MPa, whichever is greater |
Voltage | 100–240 V AC, 50/60 Hz, max. 245 W / 255 VA |