The SIL-20AC is equipped with a sample cooler with a built-in dehumidifier to minimize condensation problems. Samples can be maintained at a fixed temperature from 4ºC to 40ºC. Temperature equilibration is rapid and uniform, so heat or cold sensitive sample constituents remain stable.
Features:
- Adsorption of sample constituents has been reduced to an absolute minimum. Enhancements include:
- Special needle treatment (patent pending) – masks active sites and minimizes ionic interactions
- One piece injection port/needle seal at valve center – eliminates sample transfer tubing
- Redesigned needle seal to minimize surface area
- Low-adsorption valve materials
- These advancements combine to eliminate cross contamination – even with difficult basic and hydrophobic compounds. For especially difficult samples, dual-solvent rinsing is available.
- High Speed/Throughput – 10s for a 10 µL injection
- Ultra-high-speed sample processing – 10 seconds for a 10 µL injection – is now possible with the SIL-20A/20AC. Used together with a high-speed separation column makes an analysis cycle time of less than one minute a reality.
- Precise Sample Injection
- Greater injection volume accuracy has been attained by incorporating a high-performance precision metering pump. Injection volumes of less than 1 µL are possible, and the direct injection method means valuable samples are never wasted.
- Optional Rack Changer
- Significantly Increases Sample Capacity
- With increased speed, capacity becomes a limiting factor for unattended operation. The optional rack changer automatically loads microplates into the SIL-20A/AC sample compartment. The 12-plate capacity ensures convenient processing of over 1000 samples. The “C” model changer has the same plate capacity and incorporates a 4ºC to 40ºC cooling function, matching that of the SIL-20AC.
SIL-20AC | |
Injection method | Total-volume sample injection, variable injection volume |
Injection-volume setting range | 0.1 to 100 µL (standard), 1 to 2,000 µL (option) |
Number of processed samples | 175 (1mL vials), 70 (1.5mL vials), 50 (4mL vials), two 96-well MTPs, two 384-well MTPs, ten 1.5mL vials in addition to these |
Injection-volume accuracy | 1% max. |
Injection-volume precision | RSD: 0.3% max. (for 10 µL injection, under specified conditions) |
Cross-contamination | 0.005% max. (naphthalene, chlorhexidine) |
Number of repeated injections | 30 max. per sample |
Needle rinsing | Set freely before and after sample injection. |
Sample cooler | Block cooling/heating, used together with dehumidifying function, 4oC to 40oC |
Operating pH range | 1 – 14 |
Operating temperature range | 4oC to 35oC |
Dimensions, weight | 260 (W) x 415 (H) x 500 (D) mm, 30 kg |
Power requirements | 100 VAC, 300 VA, 50/60 Hz |