This refurbished HP / Agilent 1100 Series G1377A μ-WPS AutoSampler is in great condition. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.
| Type | Specification |
| GLP features | Early maintenance feedback (EMF), electronic records of maintenance and errors |
| Communications | Communications Controller-area network (CAN). RS232C, APG-remote standard, optional four external contact closures and BCD vial number output |
| Safety features | Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display |
| Injection range | 0.01 – 8 µl in 0.01 µl increments with the small loop capillary 0.01 – 40 µl in 0.01 µl increments with the extended loop capillary |
| Precision | Typically < 0.5 % RSD of peak areas from 5 – 40 µl, Typically < 1 % RSD from 1 – 5 µl Typically < 3 % RSD from 0.2 – 1 µl |
| Sample viscosity range | 0.2 – 5 cp |
| Sample capacity | 2 × well-plates (MTP) + 10 × 2 ml vials 100 x 2 ml in one tray 40 x 2 ml in half tray |
| Injection cycle time | Typically < 30 s using following standard conditions: Default draw speed: 4 µl/min Default eject speed: 10 µl/min Injection volume: 0.1 µl |
| Carry-over | Typically < 0.05 % using the following conditions: Column: 150 x 0.5 mm Hypersil ODS, 3 µm Mobile phase: Water/Acetonitrile = 85/15 Column Flow rate: 13 µl/min Injection volume: 1 µl caffeine (=25ng caffeine), 1 µl water to test carryover Outside wash of needle before injection: 20 sec with water using flush port |