The Agilent 7500ce ICP-MS offers trace elemental analysis which can reach down to low parts per trillion. The quadropole mass spectrometer offers the ability to scan through all the elements in a matter of seconds. The added octopole reaction system (ORS) removes interference and polyatomics. For liquid ICP-MS, the ASX-500 autosampler is available to offer automated analysis of over 200 samples. Samples can be analyzed in up to 5% acid, excluding hydrofluoric.
S/N: JP51201624
7500ce Features:
- Compact benchtop design permits easy integration into existing laboratory space
- All stainless steel chassis for superior corrosion resistance
- Open architecture sample introduction area simplifies interfacing to alternate sampling accessories
- Peltier-cooled spray chamber for optimum stability and lowest oxide interferences
- Fully automatic, computer controlled torch position in all three axes
- ShieldTorch System (STS) for highest sensitivity, best ion energy control
- Agilent Active Mass Flow Controllers (AMFC) for accurate precise control of all plasma and collision/reaction cell gases
- High efficiency digital drive 27 MHz ICP RF generator is all solid state and maintenance free
- Hyperbolic profile quadrupole rods for excellent ion transmission and superior abundance sensitivity
- Highest frequency, digitally synthesized 3 MHz RF Quadrupole Driver for enhanced performance
- Simultaneous dual mode detector with exclusive high-speed amplifier provides nine orders of linear dynamic range
- MicroMist glass concentric nebulizer
- Quartz Scott-type spray chamber
- Two Octopole Reaction System (ORS) reaction gas lines
Specifications:
| No Gas | H2 Mode | He Mode | |
|---|---|---|---|
| Sensitivity | |||
| Li (7)1 MHz/ppm | >=30 | ||
| Y (89)1 MHz/ppm | >=80 | >=30 | >=15 |
| Tl (205)1 MHz/ppm | >=40 | ||
| Background1 | <=5 CPS3 | ||
| Oxide %(CeO+)1 | <=1.5 | <=0.8 | |
| Doubly Charged % (Ce2+)1 | <=3.0 | ||
| Detection Limit ppt (3σ) | |||
| Be (9) | <=1.5 | ||
| In (115) | <=0.5 | ||
| Bi (209) | <=0.5 | ||
| Se (78) | <=4ppt | ||
| Short Term Stability (%RSD) | <=3% (20 min) | ||
| Long Term Stability (%RSD) | <=4% (2h) | ||