| Description | Vanquish Diode Array Detector FG |
| Data Collection Rate | Up to 250 Hz |
| Detection Mode | UV-vis Detection (Up to 10 Channels) |
| Detector Type | Single-Beam |
| Type | Diode Array Detector |
| Drift | < 1 mAU/hour at 254 nm |
| Flow Cell Volume | Standard, path length 10 mm (13 µL, SST) Standard, biocompatible, path length 10 mm (13 µL, PEEK) Semi-analytical, path length 7 mm (5 µL, SST) Semi-micro, pathlength 7 mm (2.5 µL, SST) Semi-micro, biocompatible, path length 7 mm (2.5 µL, titanium) Diagnostic cell |
| GLP Features | Predictive performance functions: lamp age and ignitions, lamp intensity degradation, leak detection, service monitoring period. All system parameters logged in the Chromeleon CDS Audit Trail. |
| Lamp | Deuterium lamp, tungsten lamp |
| Linearity | < 5% at 2.2 AU (typically < 5% at 2.7 AU) |
| Max. Cell Operating Pressure | Standard biocompatible: 5 MPa (50 bar, 720 psi) Other flow cells: 12 MPa (120 bar, 1740 psi) |
| Noise | <± 6 µAU at 254 nm (standard flow cell, bandwidth: 4 nm, slit width: wide, time constant: 2 sec) |
| No. of Channels | 10 + 3D field |
| PC Connection | USB 2.0; 3-port HUB to connect further Vanquish modules |
| Photodiode Array | 1024 element |
| Safety Features | Power-up diagnostics of optics, cooling fans, motors and electronics. Leak detection and safe leak handling. |
| Slit Width | Settable: Wide, narrow |
| Spectral Bandwidth | Pixel resolution: 0.6 nm (average) |
| Temperature Range | 5 to 35°C; 20 to 80% RH (non-condensing) |
| Validation Functions | Internal validation with holmium oxide filter |
| Wavelength Accuracy | ± 1 nm |
| Wavelength Range | 190–800 nm |
| Wavelength Repeatability | ± 0.1 nm |
| Weight (Metric) | 16.3 kg |
| Electrical Requirements | 100–240 V AC, 50/60 Hz, max. 245 W / 255 VA |
| Unit Size | Each |
Vanquish DAD features
- Industry leading linearity with up to 2.7 AU
- Reduced interfering refractive index and thermal effects means less baseline drift and more reliable peak integration.
- Data acquisition at rates up to 250 Hz using up to 10 absorption channels and one spectral field delivers the best support of ultrafast separations.
- Easy-to-use detector with intuitive user interaction, convenient flow cell and lamp usage tracking, and the integrated wavelength validation through holmium oxide filter.