Specifications
SPD-M20A | |
---|---|
Light source | Deuterium (D2) lamp, tungsten (W) lamp |
Number of diode elements | 512 |
Wavelength range | 190 to 800 nm |
Bandwidth, slit width | 1.2 nm (high-resolution mode), 8 nm (high-sensitivity mode) |
Wavelength accuracy | 1 nm max. |
Wavelength precision | 0.1 nm max |
Noise | 0.6 X 10-5 AU (under specified conditions) |
Drift | 5 X 10-4 AU/h (under specified conditions) |
Linearity | 2.0 AU (ASTM standard) |
Functions | Contour output, spectrum library, MAX plotting |
Cell | Optical wavelength: 10 mm, capacity: 10µL, withstand pressure: 12 MPa |
Cell temperature-control range | 5oC above room temperature to 50oC |
Web control | Parameter setting, log management, management of consumable parts, etc. |
Buffer memory | Approx. 20 minutes of data in the entire wavelength region (only when using LCsolution) |
Operating temperature range | 4oC to 35oC |
Dimensions, weight | 260 (W) x 140 (H) x 420 (D) mm, 12 kg |
Power requirements | 100 VAC, 150 VA, 50/60 Hz |
This Shimadzu SPD-M20A DAD is in great shape. It has been tested at SpectraLab Scientific.
A 90 day warranty is included.
We have special expertise on analytical equipment. We encourage you to contact us to speak with one of our expert and professional team members.