The ESA CoulArray Detector is a coulometric multi-electrode electrochemical detector for HPLC. The instrument is designed to detect and quantitate trace levels of electroactive compounds in complex matrices. A typical application is the qualitative and quantitative analysis of a broad range of biogenic amines in biological tissues.
System includes:
- DG4 Degasser
- 2x 582 Solvent Delivery Module (or 584)
- 542 Autosampler
- (optional) Column heater
- 528 UV-Vis Detecetor
- CoulArray Organizer Module
- ESA/Dionex/Thermo 5600A CoulArray coulometric multi-electrode electrochemical Detector
- PC with data
Specifications:
| Number of Electrodes: |
4, 8, 12, or 16 coulometric electrodes in series. |
| Potential Range: |
Independent potential control for each electrode from -1 volt to +2 volts in 1 mV increments, accurate to ±1 mV. |
| Current Ranges: |
±50 nA, ±5 μA, ±100 μA autoranged; full scale for each electrode and displayed from 10 pA to 100 μA. |
| Acquisition Rate: |
2 points per second on each channel. |
| Autozero: | Up to 5 μA on the 50 nA and 5 μA scale and 100 μA on the 100 μA scale. |
| Output Noise: |
<7 pA peak to peak (10 Mohm, 2 μF, low filter setting). |
| Output Resolution: |
30 fA on 50nA gain range; 3 pA on 5μA gain range and 47 pA on 100 μA gain range |
| Power Requirements: |
100/120/220/240V AC, 50/60 Hz. |
| Output Contact Closures: |
8, programmable via software. |
| Input Contact Closures: |
2 |
| Serial Communication Ports: |
RS-232 from Computer to: Pump 1, Pump 2, Autosampler, Thermal Chamber |
| Operating Temperature Range: |
10 – 35°C (instrument or control module) |
| Operating Temperature Range: |
10 – 45° C(cells) |
| Operating Humidity Range: |
40 – 80% relative humidity |
| Storage Temperature: |
10 – 60° C(instrument or control module) |
| Dimensions: | 16.75 in.(H) x 10.6 in. (W) x 17.0 in. (D) (42.5 cm x 27.0 cm x 43.0 cm). |
| Weight: | 12kg (26 lbs). |
| Certifications: | UL, CSA, CE. |