General Information:
Split/Splitless Capillary
- Range to 400 °C in 1 °C increments
- Back-pressure design permits independent adjustment of split flow rate without affecting column flow
- Septum purge built in at 3 ml/min
- Accepts columns up to 1.2-mm od
- Multimode design includes split and splitless injection
- Splitless purge time variable in 0.01-min elements
Flame Ionization Detector
- Range to 450 °C
- Grounded jet and current limited design for operator safety
- Push-button flame ignition
- Sensitive:
> 18 mCoul/gm carbon: nitrogen carrier, 0.018-in. id jet
> 15 mCoul/gm carbon: helium carrier, 0.018-in. id jet
> 22mCoul/gm carbon: nitrogen carrier, 0.018-in. id capillary jet
> 18 mCoul/gm carbon: helium carrier, 0.018-in. id capillary jet
- Minimum detectable: <5 pg carbon/sec, nitrogen carrier at S/N=2
- Linear dynamic range: <+/- 10% over a 10 7 range with 0.018-in. id jet
- Conditions: column flow 50 ml/min, 45 ml/min H 2 , 650 ml/min air, propane sample
Thermal Conductivity Detector
- Rang to 400 °C
- Single-filament (single column) design has fluidic switching of reference and analytical carrier flows
- Minimum detectable :<400 pg/ml carrier -- equivalent to <1 pmm neon in 1 ml air
- Linear dynamic range: <+/- 5% over 10 5 range
- Digital gain setting time-programmable through the keyboard Conditions: detector
100 °C, 45 ml/min switching and 30 ml/min analytical flow of helium, propane sample
Heated Devices
- Five heated zones standard: Two detectors, Two inlets, one auxiliary
- Methods stored: Two
Column Oven
- Operating range: 4 °C above ambient to 450 °C
- Setpoint entry: 1 °C for temperatures, 0.1 °C for program rates
- Programming: Rates 0.1 to 70 °C per min, 650 min maximum run time, Three ramps with initial/final holds