HPLC PRODUCTS

 

HP 6890 System


General Information:

 

Detectors: All detectors include electronic on/off of all detector gases.

    FID
  • 450° C maximum operating temperature
  • Automatic flame ignition from the keypad or ChemStation
  • Flame out detection
  • MDL: <5 pg carbon/s as propane using N2 carrier and 0.2794-mm jet
  • Linear dynamic range: < 10%, 107 with N2 carrier and 0.2794-mm jet
    NPD
  • 400° C maximum operating temperature
  • Automatic baseline adjusting via keypad or ChemStation
  • MDL: <0.4 pg N/s, <0.2 pg P/s with azobenzene/malathion mixture
  • Selectivity: 35000 to 1 gN/gC, 75000 to 1 gP/gC with azobenzene/malathion/ octadecane
  • Dynamic range: >105 N >105 P with azobenzene/malathion mixture
    ECD
  • Equipped with anode purge for contamination resistance
  • 400° C maximum operating temperature
  • Makeup gas types: argon (5%)/methane, nitrogen
  • Radioactive source: 15 mCi 63Ni plated
  • MDL: <0.04 pg/s lindane
  • Dynamic range: > 104 with lindane

 

    TCD
  • 400° C maximum operating temperature
  • Single filament with microcell
  • MDL: <400 pg propane/ml He caxrier (MDL may be affected by laboratory environment.)
  • Linear dynamic range: 105 (5%)

    Data communications

    • HP-IB, RS-232-C, and two analog output channels (1-mV, 1-V, and 10-V output available) as standard
    • INET interface optional

    Heated zones

    • Independent heated zones, not including oven: six (two inlets, two detectors, and two auxiliary)
    • Maximum operating temperatures for auxiliary zones: 400° C

    Column oven

    • Operating temperature: 4° C above ambient to 450° C
    • Temperature setpoint resolution: 1° C
    • Maximum run time: 999.99 min
    • Programming ramps/plateaus: 6/7

    Inlets

    • All inlets include septum purge.
    • Split/splitless: electronic flow, 400° C max, electronic entry of pressure or flow and split ratio version.
    • Pressure setting range: 0-100 psi
    • Total flow setting range: split/splitless, 0-200 ml/min N2, 0-1000 ml/min H2 or He

Before you make offer, please ask for real pictures and exact configurations.

We have a professional team to deal with HP line of stuffs, call us, it would make difference.

[Home] [What's New] [About Us] [Products] [Site Map] [Ordering Information] [Cooperation] [Pre-owned Equipment ] [Career]

Copyright 2003 SpectraLab Scientic Inc. All right reserved. Please contact us by email at: info@spectralabsci.com

Being supported and modified by Spectralab , contact by email at:info@spectralabsci.com