Features: - user
Friendly
- High Sensitive and selective
- Innovative electronics
and specially designed cell
| 
High
sensitivity and selectivity for automated, analytical HPLC |
User
friendly The ECD-200 Electrochemical Detector provides high sensitivity
for automated, analytical chromatography of electroactive compounds. With 0.05
to 200 nA/V detection range, this amperometric detector offers both high sensitivity
and high selectivity. The ECD-200 Electrochemical Detector is designed for
repetitive applications and increased ease of use. Two methods can be stored. The
digital display shows all important parameters for efficient interpretation: - volt
or nA/V output data
- volt or nA/V offset
- potential
- time
constraint
- base line decay
- sensitivity
- chromatogram
Additional features of the ECD-200 include: - Optimized
flow path and low dead volume of the thin layer type electrochemical cell.
- An autozero function and easy-to-use software for automated, unattended
operation.
- LCD screen that displays the real-time chromatogram and
operating parameters.
- Control by computer or through RS-232 interface
when integrated into a complete HPLC system.
Sensitive
and selective The ECD-200 provides sensitive, selective detection. The
electronic transfer between electrode and compound in solution occurs only if
the working potential is held at the right value to oxidize or reduce the compound;
the detector is blind when the compounds react above this value.
Innovative
electronics and specially designed cell
With a low dead volume, the
cell maintains efficiency of the separation. The mirror polish of all cell parts
eliminates hydrodynamic disturbances.
|
Specifications |
|
Working
Electrode |
Glassy
carbon, flat circular, inserted in lower PCTFE half-block of the flow cell |
|
Auxiliary Electrode |
Upper
stainless steel half-block | |
Reference Electrode |
Ag/AgCl/Cl-polypropylene,
screwed into auxiliary electrode | |
Spacer
Seal |
PTFE,
0.1 mm thickness | |
Cell
Volume |
2
µl from cell inlet to working electrode | |
Dummy
Cell |
Built-in, simulates electrochemical cell to test baseline noise and drift |
|
Display |
LCD
screen for real-time display of control parameters and chromatogram |
|
Software |
Keypad
entry of parameters (10 soft keys) | |
Working
Potential |
-2.55
V to +2.55 V, 10 mV increments | |
Offset
Potential |
-10
V to +10 V, 10 mV increments | |
Sensitivity
Range |
0.05
to 200 nA/V (12 steps) | |
Time
Constant |
0.5,
2, and 5 sec. | |
Manual
Control |
Autozero and event marker | |
Protocol Storage |
Two
sets of parameters can be stored | |
Noise |
Less
than 3 percent short term, peak-to-peak (following ASTM standards) |
|
Drift |
Less
than 3 % per hour at stable temperatures | |
Remote
Control |
9-pin
RS-232 connector | |
Electrical Power |
110-220
V, 50-60 Hz | |
Manufacturing Standards Safety:
EMC: |
conforms to IEC 1010-1, UL 1262 and CSA 22.2-151
conforms
to EN 50081-1 for emission and EN 50082-1 for immunity | |
Dimensions (W x D x H) |
34
x 31 x 14 cm; 13.4 x 12.2 x 5.5 inches | |
Weight |
4
kg; 8.8 lbs. | The operational
kit includes a complete cell 2 reference electrodes, fittings, gaskets. Different
working electrodes are available. Ordering information:| Part
Number | Description | | PI-105 | ECD-200
electrochemical detector |
For more information and quotations, please feel free to contact us at: info@spectralabsci.com
629 Markham Rd, P.O. Box 55552 Toronto,
ON Canada M1H 3G7
Tel: (416)438-8266, Fax: (416)438-6746 info@spectralabsci.com |