  Acquity 2996 PDA

Item

Specification

Wavelength range

190 to 800 nm

Light source

Prealigned, Deuterium lamp

Spectral Resolution/Optical Bandpass

1.2 nm per photodiode with a total of 512 photodiodes, digital and optical (3D-mode)

Digital Resolution  

1.2 nm – 600 nm (2D mode)

Wavelength accuracy

±1 nm

Linearity range

5% at 2.0 AU, propylparaben, at 256 nm

Spectral resolution

1.2 nm

Baseline noise

±1.5 × 10–5 AU peak-to-peak, dry, at 254 nm

Drift

1 × 10–3 AU/hour at 254 nm (after warm-up)

∆T  1°C per hour

Operation temperature

4 to 40 °C (39 to 104 °F), not to vary more

than 1 °C per hour (to prevent drift)

Relative humidity

20 to 80%, noncondensing

Power

Grounded ac, 100 to 240 Vac

Power requirements

100-230 VAC

Frequency range

50 to 60 Hz

Path Length

10 mm (standard analytical cell)

Cell volume

8 µl (standard analytical cell)

Pressure Limit

1000 psi (standard analytical cell)

Wetted materials

316 stainless steel, fused silica Tefzel

Dimensions

Width: 11.5 in. (29 cm)

Depth: 24 in. (61 cm)

Height: 8.5 in. (22 cm)

Weight

31.5 lbs (14.3 kg)


Water sampler manager 

Injection volume range

0.1 to 10.0 µL as standard configuration
Up to 1000.0 µL with optional extension  loop

Accuracy

+0.2 µL, measured by fluid weight removed from vial with 10.0 µL injections averaged over 20 injections using standard 100-µL syringe

Injection Linearity

>0.999 (See ACQUITY UPLC I-Class system specification document for conditions)

Maximum sample capacity

Any two of the following:
-96 and 384 microtiter plates
-48 position 2.00-mL vial plates
-48 position 0.65-mL micro-centrifuge tube plates
-24 position 1.50-mL micro-centrifuge tube plates

Sample compartment temperature range

4.0 to 40.0 °C, settable in 0.1 °C increments; maintains 19°C below ambient with a tolerance range between -2 and +4 °C

Temperature accuracy

+0.5°C at sensor

Temperature stability

+1.0°C at sensor

Injection needle wash

Integrated, active, programmable

Minimum sample required

3µL residual, using Waters’ total recovery 2-mL vials

Carryover (UV)

<0.004%
<0.001% typical

Advanced sample manager capabilities

Auto-dilution and auto-addition

Primary wetted materials

316L stainless steel, polyimide, PEEK blend, DLC, PPS


Water Binary Solvent manager 

Number of solvents

Up to four, in combination of two, A1 or A2 and B1 or B2

Solvent conditioning

Integrated vacuum degassing, six lines with two allocated for the injector needle wash/purge solvents

Gradient formation

High pressure mixing, binary gradient

Gradient profiles

11 gradient curve [ including linear, step (2), concave (4), and convex (4)]

Primary check valves

Intelligent Intake Valves

Flow accuracy

+1.0% of set flow rate at 0.500 mL/min, as per SystemsQT

Flow precision

0.075% RSD or 0.01 min SD, (0.2 to 2.0 mL/min), whichever is greater using premixed solvent

Composition ripple (baseline noise)

<1.0 mAu (See ACQUITY UPLC I-Class System Specification document for conditions)

Pump compositional precision

<0.2% RSD, or 0.02 min SD, whichever is greater (from 0.2 to 2.0 mL/min)

Pump compositional accuracy

+0.5% absolute from 5% to 95%, 0.2 to 2.0 mL/min

Compressibility compensation

Automatic, no user intervention required

Priming

Wet priming runs at a flow rate of 4 mL/min

Pump seal wash

Equipped with a programmable active wash system to flush the rear of the high pressure seals and the plungers

Flow ramping

Automatic

Primary wetted materials

316L stainless steel, UHMWPE blend, MP35N, titanium alloy, gold, sapphire, ruby, zirconia, Nitronic 60, DLC, fluoropolymer, PEEK and PEEK blend

Mixing options

Standard: 50 µL
Optional: 100 µL and 380 µL



This Waters UPLC System includes:
-Waters 2996 PDA Detector
-Waters Sample Manager
-Waters Binary Solvent manager

This refurbished Waters UPLC System is in great condition. It has been tested at SpectraLab Scientific.
 
A 90 day warranty is included.
 
We have special expertise on analytical equipment. We encourage you to contact us to speak with one of our expert and professional team members.
 
Thanks for looking and we look forward to hearing from you!

 
 
  Price: Out of Stock
  INQUIRE/OFFER
 
  Please Call Us at Toll Free Number:
1-866-777-3688 Tel: 905-946-8066
 
Contact by email at:
info@spectralabsci.com


 

The one stop shop for all your analytical laboratory needs! Your source for

PRE-OWEND Spectralab Scientific Incorporation CERTIFIED Equipment.

We buy your laboratory or surplus analytical equipment:

Contact buy@spectralabsci.com

Trade-in “service or analytical equipment”:

available via info@spectralabsci.com

*
  Feature Products:
   

Copyright 2014 SpectraLab Scientific Inc. All right reserved. Please contact us by email at: info@spectralabsci.com

Being supported and modified by Spectralab, contact by email at:info@spectralabsci.com