header

Pre-owned Equipment

Parts Depot

 

Waters ACQUITY UPLC Photodiode Array (PDA) Detector

 
Waters ACQUITY UPLC Photodiode Array (PDA) Detector
Waters ACQUITY UPLC Photodiode Array (PDA) Detector
Waters ACQUITY UPLC Photodiode Array (PDA) Detector
  The Waters ACQUITY UPLC Photodiode Array (PDA) Detector is specifically designed for Waters ACQUITY UPLC System. With data rates up to 80 Hz and noise specifications as low as 10 μAU, the ACQUITY UPLC PDA Detector offers sensitivity that is unmatched by any other photodiode array detector on the market. Further, enhanced software control provides flexibility for simultaneous 2D and 3D operation in either Empower or MassLynx Software.

Item

Specification

Wavelength range

190 to 500 nm

Optical resolution

1.2 nm

Digital resolution

1.2, 2.4, 3.6, 4.8, 6.0, 7.2, 8.4, 9.6, 10.8, 12.0

Wavelength accuracy

±1.0 nm

Wavelength Repeatability

±0.1 nm

Digital filter

Variable with data rate

Order filter

Fixed 340 nm to 500 nm

Noise (shunt in place of flow cell)

10 µAU, Peak to peak, 2 sec time constant, 30 sec

interval at 230 nm, 3.6 nm digital resolution, 2 Hz,

in a 240 micron medium shunt cell, 60 minute

warm-up time

Noise (10 mm

analytical flow cell)

14 µAU, Peak to peak, 2 sec time constant, 30 sec

interval at 230 nm, 3.6 nm digital resolution, 2 Hz,

0.5 mL/min, 10/90 Acetonitrile/water, 60 minute

warm-up time

Drift (medium shunt cell and 10 mm analytical flow cell)

1000 µAU/hour, 2 sec time constant, 30 sec

interval at 230 nm, 3.6 nm digital resolution, 2 Hz,

60 minute warm-up time.

Environmental stability: ±2 °C/hour.

Analytical flow cell conditions 0.5 mL/min, 10/90

Acetonitrile/water.

Linearity

< 5% at 2.0 AU, propylparaben series at 257 nm

Data rate

1, 2, 5, 10, 20, 40, and 80

Operating temperature

4 to 40 °C (39.2 to 104 °F)

Operating humidity

<90%, noncondensing

Shipping and storage temperature

−30 to 60 °C (−22 to 140 °F)

Shipping and storage humidity

<90%, noncondensing

Acoustic noise (instrument generated)

<65 dBA

 

This refurbished Waters ACQUITY UPLC Photodiode Array(PDA)Detector with Column Heater is in great condition. It has been testedatSpectraLab Scientific.

 

A 90 day warranty is included.

 

We are professionals. We encourage you to contact us tospeakwith one of our expert and professional team. Thanks for looking and welookforward to hearing from you!


 
 
  Price: In Stock Call for Quote
  INQUIRE/OFFER
 
  Please Call Us at Toll Free Number:
1-866-777-3688 Tel: 905-946-8066
 
Contact by email at:
info@spectralabsci.com


 

The one stop shop for all your analytical laboratory needs! Your source for

PRE-OWEND Spectralab Scientific Incorporation CERTIFIED Equipment.

We buy your laboratory or surplus analytical equipment:

Contact buy@spectralabsci.com

Trade-in “service or analytical equipment”:

available via info@spectralabsci.com

*
  Feature Products:
   

Copyright 2014 SpectraLab Scientific Inc. All right reserved. Please contact us by email at: info@spectralabsci.com

Being supported and modified by Spectralab, contact by email at:info@spectralabsci.com