General Information:


Optical performance


Wavelength range190–820nm
Slit width2 nm
Stray light <0.05 % at 220 and 340 nm, Hoya O56
Wavelength accuracy <± 2 nm NIST 2034
Wavelength reproducibility <±0.05 nm NIST 2034, ten consecutive measurements
Photometric accuracy ±0.005 A at 1 A, 440 nm, NIST 930e
Photometric noise <0.0002 A at 0 A, 500 nm, sixty 0.5-second scans, rms
Photometric stability <0.002 A/hr (at 0 A, constant ambient temp., after 1-hour warm up): measured over 1 hour, every 5 seconds
Baseline flatness <0.001 A 0.5 second blank, 0.5 second scan, rms
Typical scan time 0.5 s full range
Shortest scan time 0.1 s full range
Time until next scan 0.6 s no variance/0.1 second

Physical dimensions

  • Height × width × depth: 185 × 605 × 424 mm (7.3 × 23.8 × 16.7 inches)
  • Weight: 15.1 kg (33.2 lb)

Power requirements

  • Line voltage (selectable): 80–132 or 170–264 V AC
  • Line frequency: 47–66 Hz
  • Power consumption: 71 VA typical

Environmental conditions

  • Operating temperature: 0–40 °C (32–104 °F)
  • Non-operating temperature: -40–65 °C (-4–149 °F)

This used HP 8452 UV-Vis Spectrophotometer with Chemstation is really in a great shape and tested at SpectraLab Scientific. 90 days warranty included.


We are professional to the stuffs. Please call us.


Multiple avaialble

[Home] [What's New] [About Us] [Products] [Site Map] [Ordering Information] [Cooperation] [Pre-owned Equipment ] [Career]

Copyright 2010 SpectraLab Scientific Inc. All right reserved. Please contact us by email at: info@spectralabsci.com

Being supported and modified by Spectralab, contact by email at:info@spectralabsci.com